A novel monitoring system for a fiber Fabry-Perot interferometer (FFPI) temperature sensor has yielded a resolution of 0.013 degrees C (0.0025 fringe). Light from a broadband source passes through a scanned Michelson interferometer and is reflected from a FFPI to produce a fringe pattern, the temporal position of which is proportional to a change in the optical length of the fiber interferometer. A second Michelson interferometer with a distributed-feedback laser source is used to correct for variations in the translation rate of the motor-driven scanning mirror. Coherence multiplexing of three such sensors has also been demonstrated.