Answers are grounded in this paper and cite their sources.
Neutrons are uncharged heavy particles having the same mass as protons. Fast neutrons interact with tissue to produce mainly recoil protons from hydrogen, which contribute about 85 per cent of the absorbed dose. Neutron capture therapy, which involves flooding the tumour with boron and using slow neutrons which have very poor penetration, (Farr and Yamamoto, 1961), will not be discussed. The ionization density in the proton tracks is relatively high. The linear energy transfer (LET) for the fast neutrons from the M.R.C. cyclotron extends from about 3 to 90 keV/μ (Boag, 1954), with an average LET approximately 20 keV/μ, compared with an average of about 3 keV/μ from 250 kV X rays and less than 1 keV/μ from γ rays (Fig. 1). Fast neutron LETs depend upon the neutron energy, and therefore upon the source of the neutrons. Their values are in the region of the most rapid variation of RBE with radiation quality, as shown by Storer, Harris, Furchner and Langham (1957) and others. Biological experiments are therefore necessary with the particular neutron beam under investigation. RBE depends not only upon the LET of the radiation, but also upon the dose given and (to a smaller extent if each dose is given within a few minutes) on dose-rate.
No takes yet. Share an insight, caveat, or question.
Fowler et al. (1963) studied this question.