Answers are grounded in this paper and cite their sources.
The instrument described here is an electro-optical shutter of the Kerr cell type, and is intended for general laboratory use as a camera shutter at effective exposure times in the low millimicrosecond region. Using this shutter system, a series of photographs of exploding wires has been made at an effective exposure time of 5 mμsec. The Kerr cell has an areal aperture of 2 × 4 cm at its smaller end and an apparent angular aperture of 20°, and is hermetically sealed. The required driving voltage pulse has an amplitude of 35 kv; the synchronizing network allows the shuttering action of the Kerr cell to be synchronized with the phenomena to be studied within less than 1 mμsec. The utilization of improved switching techniques shows great promise in attaining effective exposures in the submillimicrosecond region.
No takes yet. Share an insight, caveat, or question.
Zarem et al. (1958) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: