Why the study?
QFR allows FFR calculation based on coronary angiograms using 5-F or 6-F catheters, but 4-F diagnostic catheters are routinely used in some centers and required validation.
Does Quantitative Flow Ratio (QFR) using 4-French diagnostic catheters accurately predict conventional Fractional Flow Reserve (FFR) in patients with stable coronary artery disease?
Does Quantitative Flow Ratio (QFR) using 4-French diagnostic catheters accurately predict conventional Fractional Flow Reserve (FFR) in patients with stable coronary artery disease?
QFR analysis from angiograms obtained by 4-F diagnostic catheters is feasible and shows excellent correlation with both 6-F QFR and conventional guidewire-based FFR.
OBJECTIVES: To validate QFR using 4-F diagnostic catheters compared to using 6-F guiding catheters, with conventional guidewire-based FFR as the reference standard, using independent core laboratory analysis. BACKGROUND: Quantitative Flow Ratio (QFR) allows Fractional Flow Reserve (FFR) calculation based on the coronary angiogram, using 5- or 6-French (F) catheters. However, the use of 4-F diagnostic catheters to perform coronary angiography is currently routine in some centers. METHODS: We included all consecutive patients with stable coronary artery disease and indicated for physiological assessment. QFR was performed using a 4-F diagnostic catheter, then QFR was performed using a 6-F guiding catheter while conventional FFR was measured using a pressure guidewire. Angiograms were sent to two separate core laboratories. RESULTS: One hundred lesions in 67 consecutive patients with QFR performed using 4-F and 6-F catheters, and with conventional FFR, were included. Pearson's correlation coefficient was for QFR 4-F vs. FFR 0.91 [0.87-0.94], for QFR 6-F vs. FFR 0.90 [0.86-0.94], and for QFR 4-F vs. QFR 6-F 0.93 [0.90-0.95]. Receiver-operator characteristic curves (ROC) comparing the ability to predict an FFR value above or below 0.80 with QFR 4-F and 6-F were generated. The area under the ROC curve (AUC) vs. FFR was 0.972 [0.95-0.99] for QFR 4-F and 0.970 [0.94-0.99] for QFR 6-F. CONCLUSIONS: Our study demonstrated the feasibility of performing QFR analysis from angiograms obtained by 4-F catheters, and showed a good correlation with QFR performed using 6-F catheters as well as with conventional FFR performed using a pressure guidewire.
No takes yet. Share an insight, caveat, or question.
Cuenin et al. (2021) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: