Answers are grounded in this paper and cite their sources.
Experimental imaging study demonstrates high-resolution three-dimensional tomography in a polymer blend using a single phase grating, highlighting sensitive detection of weakly absorbing materials.
X-ray phase imaging with a single phase grating based on the fractional Talbot effect is described. A phase grating with an 8 µm pitch was placed behind a weakly absorbing object and illuminated with partially coherent 17.7 keV X-rays. Intensity patterns downstream of the grating were recorded with a high-resolution image detector. By the fringe scanning method, an X-ray wavefront inclination by the object was obtained. Phase tomography was performed and the three-dimensional structure of a piece of a polymer blend was revealed with an 8 µm spatial resolution and a 9 mg/cm 3 detection limit of density deviation.
No takes yet. Share an insight, caveat, or question.
Takeda et al. (2007) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: