Answers are grounded in this paper and cite their sources.
We report a heterodyne beat with a linewidth of 5.6+/-0.6 Hz between two cavity-stabilized quantum-cascade lasers operating at 8.5 microm . We also present a technique for measuring this beat that avoids the need for extreme isolation of the optical cavities from the environment, that of employing a third servo loop with low bandwidth to force one cavity to track the slow drifts and low-frequency fluctuations of the other. Although it is not fully independent, this technique greatly facilitates heterodyne beat measurements for evaluating the performance of cavity-locked lasers above the bandwidth of the third loop.
No takes yet. Share an insight, caveat, or question.
Taubman et al. (2002) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: