A novel low-cost electronic circuit for galvanic O2 sensors enables advanced diagnostics and monitoring to improve the functional safety of life-supporting equipment.
May enhance O2 sensor safety monitoring in life-support equipment; hypothesis-generating pending prospective validation.
Self contained closed circuit rebreathers are life supporting breathing systems, where the exhaled gas is recycled by filtering CO2and replacing metabolized O2. For the O2injection control galvanic O2sensors are used to measure the partial pressure of O2. A malfunction of the sensors can lead to a gas mixture beyond life supporting limits which is life threatening. Galvanic O2sensors are often used in anaesthetic machines in hospitals. Also in this case, a malfunction of the sensors can cause severe injury to a person. It is obvious that such sensors are key elements in life supporting equipment and as such must be designed in order to achieve high functional safety. However these sensors are consuming sensors, and with that, will fail at a certain time. Failure modes include current limitation, non linearity, electrolyte evaporation or leakage, etc. Within this paper a novel low cost read out electronic circuit for galvanic O2sensors is presented, which allows not only digital readout, digital temperature compensation, on board storage of calibration and manufacturing data, but also can be used for performing voltammetry cycles to test the Pb anode for exhaustion. First measurements are presented and compared to data from a Solartron SI 1287 impedance analyzer, which is state-of-the-art in electrochemical sensor characterization.
No takes yet. Share an insight, caveat, or question.
Sieber et al. (2011) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: