Why the study?
Do Masimo and Nellcor pulse oximeters accurately measure heart rate and oxygen saturation compared to ECG in newly born term infants?
Do Masimo and Nellcor pulse oximeters accurately measure heart rate and oxygen saturation compared to ECG in newly born term infants?
Both Masimo and Nellcor pulse oximeters accurately measure heart rate in term infants, but agreement on oxygen saturation is poor at levels below 70%.
SpO2 <70% readings warrant clinical caution across devices in newborns; leaves open need for prospective device validation.
AIM: To compare heart rate (HR) measurements from Masimo and Nellcor pulse oximeters (POs) against HR measured via a three lead electrocardiograph (ECG) (HRECG ). We also compared peripheral oxygen saturation (SpO2 ) measurements between Nellcor and Masimo oximeters. METHOD: Term infants born via elective caesarean section were studied. ECG leads were placed on the infant's chest and abdomen. Masimo and Nellcor PO sensors were randomly allocated to either foot. The monitors were placed on a trolley, and data from each monitor screen captured by a video camera. HR, SpO2 measurements and signal quality were extracted. Bland-Altman analysis was used to determine agreement between HR from the ECG and each oximeter, and between SpO2 from the oximeters. RESULTS: We studied 44 infants of whom 4 were resuscitated. More than 8000 pairs of observations were used for each comparison of HR and SpO2. The mean difference (±2SD) between HRECG and HRN ellcor was -0.8 (±11) beats per minute (bpm); between HRECG and HRM asimo was 0.2 (±9) bpm. The mean (±2SD) difference between SpO2Masimo and SpO2Nellcor was -3 (±15)%. The Nellcor PO measured 20% higher than the Masimo PO at SpO2 <70%. CONCLUSION: Both oximeters accurately measure HR. There was good agreement between SpO2 measurements when SpO2 ≥70%. At lower SpO2 , agreement was poorer.
No takes yet. Share an insight, caveat, or question.
Dawson et al. (2013) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: