Why the study?
Does specific electrode placement and size optimize the current density distribution during transcutaneous cardiac pacing in a computational model?
Does specific electrode placement and size optimize the current density distribution during transcutaneous cardiac pacing in a computational model?
A finite element model demonstrates that transcutaneous pacing efficiency is optimized with a negative electrode over the cardiac apex, a positive electrode under the right scapula, and an electrode size of up to 70 cm2.
Answers are grounded in this paper and cite their sources.
May guide future transcutaneous pacing models; leaves open clinical translation and outcomes.
We developed a two-dimensional finite element model of a cross-section of the human thorax to study the current density distribution during transcutaneous cardiac pacing. The model comprises 964 nodes and 1842 elements and accounted for the electrical properties of eight different tissues or organs and also simulated the anisotropies of the intercostal muscles. The finite element software employed was a version for electrokinetics problems of Finite Element for Heat Transfer (FEHT) and we assessed the effects upon the efficacy of transcutaneous cardiac pacing of several electrode placements and sizes. To minimize pain in the chest wall and still be able to capture the heart, we minimized the ratio, R, between the current density in the thoracic wall (which causes pain) and the current density in the heart wall (which captures the heart). The best placement of the negative electrode was over the cardiac apex. The best placement of the positive electrode was under the right scapula, although other placements were nearly as good. The efficiency of pacing increased as electrode size increased up to 70 cm2 and showed little improvement for larger areas. Between different configurations of the precordial electrodes V1, V2, ..., V6 the most efficient configuration to pace with was V1 and V2 positive and V5 and V6 negative. A more efficient configuration uses an auxiliary electrode located at the right subscapular region.
No takes yet. Share an insight, caveat, or question.
Panescu et al. (1994) studied this question.