Answers are grounded in this paper and cite their sources.
For the photoelectric detection of Raman spectra excited with a He–Ne laser, the dark current of an EMI 9558 QA photomultiplier is reduced by magnetic defocusing of the electrons originating from the inefficient part of the photocathode. A lightpipe cone concentrates the light signal to a spot of 2.7 mm diameter and enhances the quantum efficiency of the photocathode by multiple reflection. A pulse counting system with a deadtime of less than 100 nsec is described and compared with dc detection. It is now possible to detect signals with a power of less than 10−17 W or less than 35 photons/sec at 650 nm with a time constant of 1 sec. The improvement in signal-to-noise ratio is demonstrated with the Raman spectrum of polycrystalline azobenzene.
No takes yet. Share an insight, caveat, or question.
Topp et al. (1969) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: