The suitability of low-cost carbon thick-film strain sensors embedded within a biomedical grade silicone rubber (Silastic® MDX4-4210) for implantable applications is investigated. These sensors address the need for robust cost-effective implantable strain sensing technology for the closed loop operation of function-restoring neural prosthetic systems. Design, fabrication and characterization of the sensors are discussed in the context of the application to strain/fullness measurements of the urinary bladder as part of the neuroprosthetic treatment of lower urinary tract dysfunction. The fabrication process, utilizing off-the-shelf screen-printing materials, is convenient and cost effective while achieving resolutions down to 75 µm. This method can also be extended to produce multilayer embedded devices by superposition of different screen-printable materials. Uniaxial loading performance, temperature dependence and long-term soak testing are used to validate suitability for implantation while proof-of-concept operation (up to 40% strain) is demonstrated on a bench-top latex balloon bladder model.