Why the study?
Does right outflow cannula banding or pump speed reduction improve pulmonary hemodynamics in a mock circulation loop using two rotary LVADs for biventricular support?
Does right outflow cannula banding or pump speed reduction improve pulmonary hemodynamics in a mock circulation loop using two rotary LVADs for biventricular support?
Both right outflow banding and pump speed reduction are effective techniques for adapting a rotary LVAD for right heart support, but require precise control due to the high sensitivity of pulmonary pressures to these parameters.
Requires precise right-pump titration due to pulmonary-pressure sensitivity; leaves open clinical translation beyond mock-loop data.
The absence of an effective, easily implantable right ventricular assist device (RVAD) significantly diminishes long-term treatment options for patients with biventricular heart failure. The implantation of a second rotary left ventricular assist device (LVAD) for right heart support is therefore being considered; however, this approach exhibits technical challenges when adapting current devices to produce the lower pressures required of the pulmonary circulation. Hemodynamic adaptation may be achieved by either reducing the rotational speed of the right pump impeller or reducing the diameter of the right outflow cannula by the placement of a restricting band; however, the optimal value and influence of changes to each parameter are not well understood. Hemodynamics were therefore investigated using different banding diameters of the right outflow cannula (3-6.5 mm) and pump speeds (500-4500 rpm), using two identical rotary blood pumps coupled to a pulsatile mock circulation loop. Reducing the speed of the right pump from 4900 rpm (for left ventricle support) to 3500 rpm, or banding the Ø10 mm (area 78.5 mm²) right outflow graft to Ø5.4 mm (22.9 mm²) produced suitable hemodynamics. Pulmonary pressures were most sensitive to banding diameters, especially when RVAD flow exceeded LVAD flow. This occurred between Ø5.3 and Ø6.5 mm (22.05-38.5 mm²) and speeds between 3200 and 4400 rpm, with the flow imbalance potentially leading to pulmonary congestion. Total flow was not affected by banding diameters and speeds below this range, and only increased slightly at higher values. Both right outflow banding or right pump speed reduction were found to be effective techniques to allow a rotary LVAD to be used directly for right heart support. However, the observed sensitivity to diameter and speed indicate that challenges may be presented when setting appropriate values for each patient, and control over these parameters is desirable.
No takes yet. Share an insight, caveat, or question.
Timms et al. (2011) studied this question.
Synapse has enriched 4 closely related papers on similar clinical questions. Consider them for comparative context: