Key result
Increasing PFA burst count from 5 to 10 increases lesion depth ~13%.
Why the study?
Pulsed field ablation therapies need to be optimized to maximize lesion depth while mitigating heat and musculoskeletal stimulation.
Absolute Event Rate: 5.3% vs 4.7%
p-value: p=<0.001
Optimization of pulsed field ablation parameters, specifically pulse width, pulses per burst, and irrigation rate, is crucial to maximize lesion depth while mitigating musculoskeletal stimulation and thermal damage.
Does not support changing clinical PFA burst counts yet; leaves open human validation of depth-safety tradeoffs.
BACKGROUND: Pulsed field ablation therapies need to be optimized to maximize lesion depth while mitigating heat and musculoskeletal stimulation. We describe the adjustment of biphasic pulsed field ablation parameters for an irrigated tip catheter. METHODS: Experimental models all used an 8F, flexible-tip, irrigated ablation catheter. Computer simulation models were designed to assess the effects of bipolar versus monopolar applications. A vegetal (potato) model was used to assess the effects of voltage, pulse width, and number of pulses per burst on lesion depth. In-vitro testing on bovine cardiac muscle was performed with temperature sensors at the surface and 3 and 7 mm beneath the surface to measure temperature rises. Finally, an in vivo canine model was used to assess for musculoskeletal stimulation caused by the pulsed field ablation therapies established with the vegetal/in vitro testing. RESULTS: Monopolar delivery consistently delivered more tissue depth than bipolar configurations. Adjustments from n to n+1.0 µs and increasing pulses per burst from 50 to 250 increased lesion depth but also increased musculoskeletal stimulation. The optimized burst count was 5 R-wave-gated bursts over 4 to 9 seconds. The optimized therapy had a minimal observed depth of >5 mm and minimized musculoskeletal stimulation. Final selected therapies did create temperature rises, which were completely offset by irrigation rates of 7 to 13 mL/min. CONCLUSIONS: Pulse width and number of pulses per burst are important contributors to lesion depth, musculoskeletal stimulation, and incident heating. Even small adjustments in these parameters can alter lesion depth, muscle stimulation, and thermal effects, which will differentiate optimal from suboptimal pulsed field ablation.
No takes yet. Share an insight, caveat, or question.
Verma et al. (2026) studied Cardiac ablation (n=3). Pulsed field ablation parameter adjustment vs. 5 bursts over 1.0 sec was evaluated on Mean lesion depth (p=<0.001). Increasing pulsed field ablation burst count from 5 to 10 significantly increased mean lesion depth from 4.7 mm to 5.3 mm, though optimized parameters are required to balance depth with musculoskeletal stimulation.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: