Key result
The devised automated oscillometric NIBP monitor significantly underestimated systolic blood pressure by a mean of 7.38 mmHg compared to a standard mercury sphygmomanometer (119.65 vs 127.03 mmHg, p=0.0013).
Why the study?
Does a devised automated oscillometric NIBP monitor provide accurate blood pressure measurements compared to a standard mercury sphygmomanometer in healthy young adults?
Cross-Sectional (n=30)
No
Does a devised automated oscillometric NIBP monitor provide accurate blood pressure measurements compared to a standard mercury sphygmomanometer in healthy young adults?
Mean Difference: -7.38
Absolute Event Rate: 119.65% vs 127.03%
p-value: p=0.0013
The devised automated oscillometric NIBP monitor consistently underestimates blood pressure compared to a mercury sphygmomanometer but demonstrates high correlation, making it suitable for educational and training purposes.
Should not guide clinical decisions due to SBP underestimation; leaves open validation for educational use.
Objectives: This paper focuses on the evaluation of accuracy and reliability of previously designed oscillometric based Non Invasive Blood Pressure (NIBP) instrument by means of a comparative analysis with the standard cuff mercury sphygmomanometer based on auscultation. Methods/Statistical Analysis: Thirty healthy subjects between the ages of 18-23 were examined by successive application of both devices having identical cuffs and site of measurement. The analysis of means and bias values demonstrated that our designed oscillometric NIBP monitor under-read all blood pressure values compared to mercury manometer. Findings: The mean oscillometric NIBP readings (Systolic Blood Pressure (SBP) 119.65 ± 8.25, Diastolic Blood Pressure (DBP) 76.82 ± 6.12 and Mean Arterial Pressure (MAP) 91.1 ± 4.76) were significantly lower than its auscultatory counterpart (SBP 127.03 ± 8.7, DBP 81.4 ± 6.2 and MAP 96.6 ± 4.95). F-test showed that the devised prototype is precise and comparable to a reference mercury manometer. A p-value of <0.05 was recognized to be statistically considerable. Conclusively, the designed NIBP monitoring prototype exhibited a tendency to measure lower blood pressure values than the auscultatory sphygmomanometer; but both devices were well comparable. The difference found, was statistically significant but below the level of physiological significance. Application/Improvements: This oscillometric NIBP prototype may be used for training as well as learning objectives. Furthermore, this prototype can be improved by Arduino based microcontroller and kit in the future studies.
No takes yet. Share an insight, caveat, or question.
Shahid et al. (2017) conducted a cross-sectional in Healthy (n=30). Devised automated oscillometric NIBP monitor vs. Standard mercury sphygmomanometer was evaluated on Systolic Blood Pressure (SBP) (MD -7.38, p=0.0013). The devised automated oscillometric NIBP monitor significantly underestimated systolic blood pressure by a mean of 7.38 mmHg compared to a standard mercury sphygmomanometer (119.65 vs 127.03 mmHg, p=0.0013).