Answers are grounded in this paper and cite their sources.
Manganin gauges were calibrated at five fixed points, and the resulting calibration compared with that obtained at a single fixed point, or at three fixed points. For most work three points are sufficient. The nonlinearity is 0.135%/kbar. Manganin gauges intercompared agree within one bar to 10 kbar. Pressures of the H2O L-III-V, L-V-VI triple points and melting pressure of VI at 25°C were determined. It was not found possible to prepare a gold-chromium coil which was free of hysteresis.
No takes yet. Share an insight, caveat, or question.
Boren et al. (1965) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: