A method is proposed in which the responsivity (A/Im) for the photometric quantity of a silicon photodiode, which is combined with a V(lambda) filter of a known spectral transmittance, can be directly self-calibrated by measuring the surface reflectance and the recombination losses of the photodiode for white light. The derived equation for this method assumes approximations that can cause systematic errors. A computer simulation was made to quantify these errors, and an accuracy of better than 0.2% was predicted. Although experimental validation is necessary, this method is promising as a new, practical method for high-accuracy photometric standards.
No takes yet. Share an insight, caveat, or question.
Yoshihiro Ohno (1992) studied this question.