PURPOSE: To evaluate the diagnostic accuracy of a nonenhanced electrocardiograph-gated quiescent-interval single shot MR-angiography (QISS-MRA) at 3 Tesla with contrast-enhanced MRA (CE-MRA) and digital subtraction angiography (DSA) serving as reference standard. MATERIALS AND METHODS: Following institutional review board approval, 16 consecutive patients with peripheral arterial disease underwent a combined peripheral MRA protocol consisting of a large field-of-view QISS-MRA, continuous table movement MRA, and an additional time-resolved MRA of the calves. DSA correlation was available in eight patients. Image quality and degree of stenosis was assessed. Sensitivity and specificity of QISS-MRA was evaluated with CE-MRA and DSA serving as the standards of reference and compared using the Fisher exact test. RESULTS: With the exception of the calf station, image quality with QISS-MRA was rated statistically significantly less than that of CE-MRA (P < 0.05, P = 0.17, and P = 0.6, respectively). A greater percentage of segments were not accessible with QISS-MRA (19.5-20.1%) in comparison to CE-MRA (10.9%). Relative to DSA, sensitivity for QISS-MRA was high (100% versus 91.2% for CE-MRA, P = 0.24) in the evaluated segments; however, specificity (76.5%) was substantially less than that of CE-MRA (94.6%, P = 0.003). CONCLUSION: Overall image quality and specificity of QISS-MRA at 3T are diminished relative to CE-MRA. However, when image quality is adequate, QISS-MRA has high sensitivity and, thus, has potential use in patients with contraindications to gadolinium.
No takes yet. Share an insight, caveat, or question.
Hansmann et al. (2013) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: