A system design and automation of a microbiological instrument that locates bacterial colonies and captures the forward-scattering signatures are presented. The proposed instrument integrates three major components: a colony locator, a forward scatterometer and a motion controller. The colony locator utilizes an off-axis light source to illuminate a Petri dish and an IEEE1394 camera to capture the diffusively scattered light to provide the number of bacterial colonies and two-dimensional coordinate information of the bacterial colonies with the help of a segmentation algorithm with region-growing. Then the Petri dish is automatically aligned with the respective centroid coordinate with a trajectory optimization method, such as the Traveling Salesman Algorithm. The forward scatterometer automatically computes the scattered laser beam from a monochromatic image sensor via quadrant intensity balancing and quantitatively determines the centeredness of the forward-scattering pattern. The final scattering signatures are stored to be analyzed to provide rapid identification and classification of the bacterial samples.