Radiation therapy can cause malfunction in pacemakers, particularly newer multiprogrammable units, due to radiation-induced damage to integrated circuits.
Answers are grounded in this paper and cite their sources.
Radiation may damage pacemaker circuits during therapy; case report leaves open incidence and shielding needs.
Pacemaker malfunction occurred after a patient was given 3000–3600 rads to an area occupied by an A‐V sequential pacemaker. Analysis of the removed generator showed that there was malfunction of the large scale integrated circuit and the type of damage was consistent with radiation‐induced effects. The newer muJtiprogram‐mable units may be more sensitive to ionizing radiation than those previously available. (PACE, Vol. 5, March‐April, 1982)
No takes yet. Share an insight, caveat, or question.
Katzenberg et al. (1982) studied this question.