Why the study?
Is electrical stimulation safe from causing electromagnetic interference in patients with pacemakers and implantable cardioverter defibrillators?
Is electrical stimulation safe from causing electromagnetic interference in patients with pacemakers and implantable cardioverter defibrillators?
Electrical stimulation of the lower limb appears relatively safe in patients with cardiac devices, but caution is advised until larger safety studies are conducted.
Answers are grounded in this paper and cite their sources.
Supports cautious use of lower-limb stimulation in cardiac device patients; leaves open precise risk quantification.
INTRODUCTION: A number of patients are excluded from electrical stimulation treatment because there is concern that electrical stimulation could cause electromagnetic interference with pacemakers and implanted cardioverter defibrillators. The decision to use electrical stimulation in these patients needs to be supported by an assessment of benefit and harm. METHODS: We conducted a systematic review of the risk of electromagnetic interference between electrical stimulation and pacemakers or implanted cardioverter defibrillators. We included the electronic databases MEDLINE and EMBASE in the time period between 1966 and 26 August 2016. RESULTS: 18 papers fulfilled the inclusion criteria (eight safety studies and ten case studies). Although we were unable to accurately estimate the risk of electromagnetic interference, the studies revealed that patients having electrical stimulation of the lower limb are less susceptible to electromagnetic interference. CONCLUSIONS: The results suggest that electrical stimulation could be used safely to help drop foot in patients with pacemakers or implanted cardioverter defibrillators. However, in order to obtain an accurate estimate of the risk of electromagnetic interference, a large, long-term, and intervention-specific safety study is required. Until such a study is undertaken, electrical stimulation should be used with caution in patients with pacemakers and implanted cardioverter defibrillators.
No takes yet. Share an insight, caveat, or question.
Badger et al. (2017) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: