Why the study?
What are the trends, indications, and outcomes of cardiac implantable device system extraction in a high-volume centre over a decade?
What are the trends, indications, and outcomes of cardiac implantable device system extraction in a high-volume centre over a decade?
Percutaneous lead extraction in a high-volume centre is associated with a high procedural success rate (98.9%) and low major complication rate (1%), with duration of lead implantation being the only indicator for success.
Answers are grounded in this paper and cite their sources.
Supports extraction feasibility in high-volume centers; leaves open generalizability and need for prospective validation.
BACKGROUND: The rising number of device implantation has seen a parallel in the rising numbers of lead extraction. Herein we have analysed our experience in cardiac device and lead extraction in a single tertiary centre over the last decade. METHOD: Retrospective analysis of all consecutive patients undergoing lead extractions performed between 2001 and 2010. Procedural success and complications as defined by the Heart Rhythm Society policy. RESULTS: A total of 745 leads were extracted with a procedural success of 98.9% [382 cases; partial success in 6.9% (26) cases] and failure in 1.1% (4). Major complication rate was 1% (four cases) and minor complication rate was 3.6%. By both univariate and multivariate analysis only duration of lead implantation was an indicator for success (p < 0.0001). The mean implantation time for failed lead extraction was 203 ± 64 months compared with 71.8 ± 16.5 months in the successful cohort (p < 0.0001). Laser-assisted extraction was required in 176 cases. With regard to extraction indication, lead malfunction/recall showed a significant increase during the study period (p = 0.03). On time trend analysis the rise in coronary sinus (CS) lead extraction over time was significant. (p = 0.02) Despite a trend for increased laser use over time this did not achieve statistical significance, p = 0.06. CONCLUSIONS: A decade's experience of percutaneous lead extraction suggests that a high procedural success rate with a low complication rate is achieved in a high-volume centre. During this time, an increase in both defibrillator and CS lead explantation and a rising trend in laser assistance with almost 50% of cases needing laser usage were observed.
No takes yet. Share an insight, caveat, or question.
Arujuna et al. (2012) studied this question.