OBJECTIVE: To evaluate a manual method (Cytosphere) for quantifying CD4+ T-cell numbers. DESIGN: Cross-sectional study of HIV-1-seronegative and HIV-1-seropositive individuals evaluated for absolute CD4 counts by both standardized flow cytometric measurements and manual Cytosphere technology using a hemacytometer. SETTING: University research hospitals in both the United States and Africa. PATIENTS, PARTICIPANTS: Blood specimens from 382 patients were evaluated. These were broken down into 294 samples obtained from HIV-1-seropositive patients and 88 samples obtained from HIV-1-seronegative patients. INTERVENTIONS: None. OUTCOME MEASURED: Absolute CD4 cell number. RESULTS: Evaluation of samples obtained from HIV-1 patients in both the United States and Africa demonstrated an overall correlation of the Cytosphere assay with flow cytometry of 0.912 (95% confidence interval, 0.895-0.928; P < 0.001). When samples were stratified based on CD4+ T-cell counts determined by flow cytometry, the Cytosphere assay had a 96% predictive value for correctly identifying individuals with CD4 T-cell counts > 200 x 10(6)/l and a 92% predictive value for correctly identifying individuals with CD4 T-cell counts < 200 x 10(6)/l. CONCLUSIONS: This assay appears to have the potential for the quantitation of CD4 cells in the limited laboratory facilities in developing countries and to have a strong correlation with standard flow cytometric technology.
No takes yet. Share an insight, caveat, or question.
Landay et al. (1993) studied this question.