Answers are grounded in this paper and cite their sources.
An electro-deposition approach was established to incorporate the gold nano-flakes onto the glassy carbon electrode in electrochemical cells (nano-Au/GC/ECCs). Using pulsed amperometric detection (PAD) without any gold oxidation for cleaning (non-oxidative PAD), the nano-Au/GC/ECCs were able to maintain their activity for oxidizing of carbohydrates in a normal alkaline medium. The reproducibility of peak area was about 2 relative standard deviation (RSD,%) for 6 consecutive injections. A dynamic range of carbohydrates was obtained over a concentration range of 5–80 mg L −1 and the limits of detection (LOD) were of 2 mg L −1 for fructose and lactose and 1 mg L −1 for glucose and galactose. Moreover, the nano-Au/GC/ECC using the non-oxidative PAD was able to combine with the internal standard method for determination of lactose in fresh cow milk sample.
No takes yet. Share an insight, caveat, or question.
Nguyen et al. (2015) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: