This study presents a cost-effective automation solution for preparing polymerase chain reaction (PCR) reagent cartridges used in automated nucleic acid analyzers. Manual preparation is labor-intensive and error-prone, often causing inaccurate volumes and reagent mismatches. To address this, a dispensing system based on open-source 3D printer technology is developed. It incorporates a motion platform and a syringe-based pump, and precisely dispenses reagents into cartridge chambers designed for magnetic DNA extraction and real-time PCR. The system is evaluated for manual inefficiency and error. Dispensing accuracy, assessed gravimetrically using 500 μL of distilled water, shows a relative accuracy of 0.30% and a coefficient of variation (CV) of 2.64%, both within ISO 8655 limits. In terms of efficiency, the system fills a single cartridge chamber in 13.57 seconds, much faster than the approximately 3 minutes required for manual reagent injection. These results highlight the system’s potential to improve throughput and precision in cartridge preparation.