Why the study?
Unibody bifurcated endografts can reduce operative time and prevent migration and iliac limb dislocation, prompting evaluation of long-term experience with the Endologix AFX system.
What are the long-term outcomes of endovascular aortic repair using the Endologix AFX unibody bifurcated endograft in patients with infrarenal abdominal aortic aneurysm?
What are the long-term outcomes of endovascular aortic repair using the Endologix AFX unibody bifurcated endograft in patients with infrarenal abdominal aortic aneurysm?
The Endologix AFX unibody bifurcated endograft provides safe early outcomes for infrarenal AAA, but requires close long-term monitoring due to a 17.6% endoleak rate, particularly in aneurysms larger than 6 cm.
AFX endograft feasible for infrarenal AAA over 3+ years; leaves open need for randomized comparisons to other devices.
Unibody bifurcated endografts have the advantage of reducing the operative time, avoiding migration and iliac limb dislocation in patient with abdominal aortic aneurysm (AAA). We report our long-term experience in patients who underwent endovascular aortic repair (EVAR) due to infrarenal AAA with Endologix AFX® endograft system. Between January 2013-December 2018, 68 patients with infrarenal AAA had EVAR procedure with Endologix AFX® endograft system. Mean follow-up was 40.4 ± 19.5 months, and all patients had computed tomography periodically, with colored Doppler ultrasonography (CDUS) every six months. Mean age was 68.5 ± 7.1 years and, 63 (92.6%) patients were male. Early mortality, renal complications, stent-graft migration and cardiac complications were not seen in early post-operative period. There was no early mortality in the group and no conversion to open repair. In long-term follow-up 12 patients (17.6%) had endoleak (5 with type II, 7 with type III). Overall survival estimated by Kaplan-Meier analysis was 94.1% at 1 year, 85.2% at 2 years, 74.1% at 3 years and 54.0% at 5 years. Freedom from second intervention and conversion was 98.4% at 1 year, 95.3% at 2 years, 93.3% at 3 years and 87.4% at 5 years. Endologix AFX® endograft provides a fast, safe and effective endograft for the early period, therefore it seems more suitable for elderly and comorbid patients. This endograft has low reintervention rates however according to our results, aneurysms larger than 6 cm may have more sideway displacement possibility and by this way type III endoleak. Proper patient selection and sufficient overlap are the key issues. Close monitoring is mandatory at the follow-up period.
No takes yet. Share an insight, caveat, or question.
İşcan et al. (2021) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: