The background conditions achieved in a simple ultrahigh-vacuum system with an envelope of aluminosilicate glass (Corning 1720) have been studied. Pressures were measured by a dynamic method and also with a modulated Bayard–Alpert gauge. Good agreement between the two methods was obtained using a modified analysis for the modulated gauge. A total pressure of 8 × 10−13 Torr (equivalent nitrogen) was measured with the whole system at room temperature, and one of 7 × 10−15 Torr (equivalent nitrogen) was measured with a portion of the system immersed in liquid helium. A discussion is given of variations in the residual modulation factor of the gauge in the presence of extremely low gas pressures.
No takes yet. Share an insight, caveat, or question.
J. P. Hobson (1964) studied this question.