Key result
Novel balloon-assisted modified shoe-horn technique enables successful transcatheter aortic valve delivery in horizontal bicuspid anatomy.
Why the study?
TAVI in bicuspid aortic valve poses unique challenges, and apparently benign anatomies can lead to failure of conventional delivery strategies.
Does a modified balloon-assisted shoe-horn technique enable successful TAVI delivery in patients with functionally horizontal anatomy due to bicuspid aortic valve?
Case Report (n=1)
Does a modified balloon-assisted shoe-horn technique enable successful TAVI delivery in patients with functionally horizontal anatomy due to bicuspid aortic valve?
A modified balloon-assisted shoe-horn technique can facilitate TAVI valve delivery in bicuspid aortic valve patients with functionally horizontal anatomy when conventional methods fail.
Background Transcatheter aortic valve implantation (TAVI) in bicuspid aortic valve (BAV) poses unique challenges. We report a case in which an apparently benign anatomy resulted in failure of conventional delivery strategies and required a novel balloon-assisted bailout technique. Case summary A 59-year-old man with a background history of liver cirrhosis presented with severe aortic stenosis due to type 1 BAV. The patient was in Child-Pugh class A with MELD score of 14, hence he was planned for TAVI in view of high surgical risk. Pre-procedural computed tomography showed a favorable LV-Aortic angle, but the fused non-coronary and right-coronary cusp behaved like functional horizontal anatomy. Various previously described wire and balloon-assisted techniques failed to deliver the valve. A novel balloon-assisted modified “shoe-horn” technique, by inflating a 16 mm balloon in non-coronary sinus, altered the delivery trajectory and enabled successful valve delivery. Discussion BAV with right-noncoronary cusps fusion may behave like horizontal anatomy, despite favorable LV-Aortic angle. A modified balloon-assisted shoe-horn technique can facilitate valve delivery when conventional methods fail. This case highlights an underrecognized mechanism of valve delivery failure in BAV and emphasizes the importance of preparedness for advanced bailout strategies.
No takes yet. Share an insight, caveat, or question.
Patil et al. (2026) conducted a case report in Severe aortic stenosis due to type 1 bicuspid aortic valve (n=1). Modified balloon-assisted shoehorn technique was evaluated on Successful valve delivery. A novel balloon-assisted modified shoe-horn technique enabled successful transcatheter aortic valve delivery in a patient with functionally horizontal bicuspid aortic valve anatomy.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: