Answers are grounded in this paper and cite their sources.
This proof of concept study presents a method to collect and analyse kinetic data from one participant with a transfemoral amputation fitted with a percutaneous osseointegrated implant walking on a level and sloped treadmill. We describe the construction of and results from a bespoke wireless six axis load cell built into one participant's prosthetic assembly. The load cell does not clinically compromise the participant in any way and is an initial milestone in the development of a light-weight wireless load cell for use with percutaneous osseointegrated implants. In this case, it is the first time that kinetic data from a participant fitted with an Intraosseous Transcutaneous Amputation Prosthesis has been published. We propose that the data can be used to model the load transfer to the host bone, with several clinically significant applications. The raw dynamic data are made available and quasi-static load cases for each functional phase of gait are presented. Peak forces obtained in the medio-lateral (X), cranio-caudal (Y) and antero-posterior (Z) axes over level ground respectively were -243.8 N (0.24 BW), 1321.5 N (1.31 BW) and -421.8 N (0.42 BW); uphill were -141.0 N (0.14 BW), 1604.2 N (1.59 BW), -498.1 N (0.49 BW); downhill were -206.0 N (0.20 BW), 1103.9 N (1.09 BW), -547.2 N (0.54 BW). The kinetics broadly followed able bodied gait patterns with some gait strategies consistent in participants with other implant designs or prosthetic socket connections, for example offloading the artificial limb downhill.
No takes yet. Share an insight, caveat, or question.
Ahmed et al. (2024) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: