Key result
A model-based intelligent active relative motion canceling algorithm using biological signals was developed and tested on a 3-degree-of-freedom robotic test-bed system for off-pump CABG surgery.
Answers are grounded in this paper and cite their sources.
A novel model-based intelligent active relative motion canceling algorithm using ECG signals shows promise for dynamically stabilizing the heart during robotic-assisted off-pump CABG surgery.
This paper focuses on the development of control algorithms for intelligent robotic tools that assist off-pump coronary artery bypass graft (CABG) surgery. In the robotic-assisted CABG surgery, the surgeon operates on the beating heart using intelligent robotic instruments. Robotic tools actively cancel the relative motion between the surgical instruments and the point of interest on the beating heart, dynamically stabilizing the heart for the operation. This algorithm is called active relative motion canceling (ARMC). Here, a model-based intelligent ARMC algorithm employing biological signals, such as electrocardiogram, to achieve effective motion cancellation is proposed. Finally, experimental results of the algorithm on a 3-degree-of-freedom robotic test-bed system are reported.
No takes yet. Share an insight, caveat, or question.
Bebek et al. (2007) studied off-pump coronary artery bypass graft (CABG) surgery. model-based intelligent active relative motion canceling (ARMC) algorithm was evaluated on effective motion cancellation on a 3-degree-of-freedom robotic test-bed system. A model-based intelligent active relative motion canceling algorithm using biological signals was developed and tested on a 3-degree-of-freedom robotic test-bed system for off-pump CABG surgery.
Synapse has enriched 10 related papers. Consider them for comparative context: