Answers are grounded in this paper and cite their sources.
Background: The thermal effect correlated with implant osteotomy could produce significant effects on the healing process and fixture osseointegration. The aim of the present investigation was to assess the heat generation and surface wearing of dental implant drills and manual tappers during simulated osteotomies on animal ribs. Methods: Steel drills (20 units per type) and tappers (20 units per type) were evaluated for a total of 30 osteotomies. The infrared thermal analysis was performed at the first and thirtieth osteotomy. The surface alteration and wearing was assessed by energy dispersive spectroscopy–scanning electron microscopy (EDS-SEM) prior to and after use. Conclusions: The drill material produced a non-significant temperature change during bone osteotomy. Lower heating was reported for manual tappers in favor of a manual osteotomy instead rotary instruments.
No takes yet. Share an insight, caveat, or question.
Lorusso et al. (2022) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: