Key result
CentriMag circulation drives ~10-fold higher platelet GPIIb/IIIa activation alongside reduced surface receptor expression.
Why the study?
Mechanical circulatory support devices can damage blood and platelets, but the device-induced alterations of major platelet surface receptors, VWF, and associated hemostatic functions remained to be investigated.
Does continuous circulation through a CentriMag rotary blood pump for 4 hours alter platelet surface receptors, von Willebrand factor multimers, and hemostatic adhesion functions in human whole blood?
Does continuous circulation through a CentriMag rotary blood pump for 4 hours alter platelet surface receptors, von Willebrand factor multimers, and hemostatic adhesion functions in human whole blood?
Rotary blood pumps simultaneously induce platelet GPIIb/IIIa activation (promoting fibrinogen adhesion and thrombosis) and progressive shedding of GPVI and GPIbα receptors alongside rapid HMWM-VWF loss (impairing collagen/VWF adhesion and promoting bleeding), offering a mechanistic explanation for concurrent thrombotic and bleeding risks in mechanical circulatory support.
Answers are grounded in this paper and cite their sources.
Indicates dual thrombotic-bleeding platelet effects during rotary support; leaves open clinical outcome prediction and in vivo validation.
Thrombotic and bleeding complications are the major obstacles for expanding mechanical circulatory support (MCS) beyond the current use. While providing the needed hemodynamic support, those devices can induce damage to blood, particularly to platelets. In this study, we investigated device-induced alteration of three major platelet surface receptors, von Willebrand factor (VWF) and associated hemostatic functions relevant to thrombosis and bleeding. Fresh human whole blood was circulated in an extracorporeal circuit with a clinical rotary blood pump (CentriMag, Abbott, Chicago, IL, USA) under the clinically relevant operating condition for 4 hours. Blood samples were examined every hour for glycoprotein (GP) IIb/IIIa activation and receptor loss of GPVI and GPIbα on the platelet surface with flow cytometry. Soluble P-selectin in hourly collected blood samples was measured by enzyme linked immunosorbent assay to characterize platelet activation. Adhesion of device-injured platelets to fibrinogen, collagen, and VWF was quantified with fluorescent microscopy. Device-induced damage to VWF was characterized with western blotting. The CentriMag blood pump induced progressive platelet activation with blood circulating time. Particularly, GPIIb/IIIa activation increased from 1.1% (Base) to 11% (4 hours) and soluble P-selectin concentration increased from 14.1 ng/mL (Base) to 26.5 ng/mL (4 hours). Those device-activated platelets exhibited increased adhesion capacity to fibrinogen. Concurrently, the CentriMag blood pump caused progressive platelet receptor loss (GPVI and GPIbα) with blood circulating time. Specifically, MFI of the GPVI and GPIbα receptors decreased by 17.2% and 16.1% for the 4-hours sample compared to the baseline samples, respectively. The device-injured platelets exhibited reduced adhesion capacities to collagen and VWF. The high molecular weight multimers (HMWM) of VWF in the blood disappeared within the first hour of the circulation. Thereafter the multimeric patterns of VWF were stable. The change in the VWF multimeric pattern was different from the progressive structural and functional changes of platelets with the circulation time. This study suggested that the CentriMag blood pump could induce two opposite effects on platelets and associated hemostatic functions under a clinically relevant operating condition. The device-altered hemostatic function may contribute to thrombosis and bleeding simultaneously as occurring in patients supported by a rotary blood pump. Device-induced damage of platelets may be an important cause for bleeding in patients supported with rotary blood pump MCS systems relative to device-induced loss of HMWM-VWF.
No takes yet. Share an insight, caveat, or question.
Chen et al. (2019) studied Device-induced platelet dysfunction and hemostatic alterations in mechanical circulatory support (n=7). CentriMag rotary blood pump extracorporeal circulation vs. Baseline blood sample collected prior to CentriMag blood pump operation was evaluated on Platelet GPIIb/IIIa activation at 4 hours (p=<0.05). CentriMag circulation for 4 hours increased platelet GPIIb/IIIa activation from 1.1% to 11% while decreasing GPVI and GPIbα receptor expression by 17.2% and 16.1%.