Answers are grounded in this paper and cite their sources.
Spectral computed tomography (CT) systems are employed with energy-resolving photon counting detectors. Incorporation of a spectrally accurate x-ray beam model in image reconstruction helps to improve material identification and quantification by these systems. Using an inaccurate x-ray model in spectral reconstruction can lead to severe image artifacts, one of the extreme cases of this is the well-known beam-hardening artifacts. An often overlooked spectral feature of x-ray beams in spectral reconstruction models is the angular dependence of the spectrum with reference to the central beam axis. To address these factors, we have developed a parameterized semi-analytical x-ray source model in the diagnostic imaging range (30-120 kVp) by applying regression techniques to data obtained from Monte Carlo simulations (EGSnrc). This x-ray beam model is generalized to describe the off-axis spectral information within ±17 o along θ (vertical direction), ±5 o along ϕ (horizontal direction) of the central axis, and can be parameterized for specific x-ray tube models. Comparisons of our model with those generated by SpekCalc, TOPAS, and IPEM78 at central axis show good agreement (within 2 %). We have evaluated the model with experimental data collected with a small animal spectral scanner.
No takes yet. Share an insight, caveat, or question.
Shamshad et al. (2017) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: