Answers are grounded in this paper and cite their sources.
Silicon surface barrier diode detectors of 1 cm2and 25 mm2sensitive area have given pulse height spectral resolutions of 17 kev and 13-½ kev (FWHM) respectively, for 5.5 mev alpha particles. Reverse currents at 500 volts bias were less than 1 × 10-6amps/cm2with breakdown in excess of 1000 volts. A charge sensitive amplifier contributes 3-½ and 10 kev noise (FWHM) with an input capacitive loading of 20 and 180 pf respectively.
No takes yet. Share an insight, caveat, or question.
Blankenship et al. (1961) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: