Answers are grounded in this paper and cite their sources.
BACKGROUND: Urine output is a surrogate for tissue perfusion and is typically measured at 1-hour intervals. Because small urine volumes are difficult to measure in urine collection bags, considerable over- or underestimation is common. To overcome these shortcomings, digital urine meters were developed. Because these monitors measure urine volume in 1-minute intervals, they provide minute-to-minute measurements of the urine flow rate (UFR). In a previous study, we observed that the minute-to-minute variability in the UFR disappeared during hypovolemia. The aim of this study was to describe the minute-to-minute variability in the UFR as a new physiological variable and to show its relationship to blood volume depletion. METHODS: Seven adult pigs were used in this study. The UFR, minute-to-minute UFR, mean arterial blood pressure, heart rate, and base excess were measured at euvolemia and during gradual hemorrhaging (10%, 20%, and 30% of estimated blood volume). Variance and wavelet spectral analysis were used to measure the disappearance of the minute-to-minute UFR variability. RESULTS: The UFR decreased from 2.2 ± 0.2 to 1.0 ± 0.1 mL/min after a 10% estimated blood volume loss (±1 SE, n = 7, P = 0.0348). The variance in the minute-to-minute UFR decreased from 1.4 ± 0.3 to 0.4 ± 0.1 mL/min (±1 SE, n = 7, P = 0.046). CONCLUSIONS: The UFR and its minute-to-minute variability decrease during hemorrhaging. The variability in the UFR may be useful as an aid for the diagnosis of hypovolemia.
No takes yet. Share an insight, caveat, or question.
Klein et al. (2012) studied this question.