Why the study?
LVAD-induced hemodynamics feature fast-moving flow with large velocity variations, making quantitative assessments difficult with existing imaging modalities.
Does 1,000 fps high-speed angiography accurately quantify the hemodynamic effects of different LVAD outflow graft angles compared to CFD simulations in vitro?
Does 1,000 fps high-speed angiography accurately quantify the hemodynamic effects of different LVAD outflow graft angles compared to CFD simulations in vitro?
High-speed angiography at 1,000 fps is a viable quantitative imaging modality for evaluating fast-moving hemodynamics in LVAD outflow grafts in vitro.
Supports HSA for quantitative in vitro LVAD graft flow assessment; leaves open in vivo validation before clinical use.
Left ventricular assist device (LVAD)-induced hemodynamics are characterized by fast-moving flow with large variations in velocity, making quantitative assessments difficult with existing imaging methods. This study demonstrates the ability of 1,000 fps high-speed angiography (HSA) to quantify the effect of the surgical implantation angle of a LVAD outflow graft on the hemodynamics within the ascending aorta in vitro . High-speed angiography was performed on patient-derived, three-dimensional-printed optically opaque aortic models using a nonsoluble contrast media, ethiodol, as a flow tracer. Outflow graft configuration angles of 45° and 90° with respect to the central aortic axis were considered. Projected velocity distributions were calculated from the high-speed experimental sequences using two methods: a physics-based optical flow algorithm and tracking of radio-opaque particles. Particle trajectories were also used to evaluate accumulated shear stress. Results were then compared with computational fluid dynamics (CFD) simulations to confirm the results of the high-speed imaging method. Flow patterns derived from HSA coincided with the impingement regions and recirculation zones formed in the aortic root as seen in the CFD for both graft configurations. Compared with the 45° graft, the 90° configuration resulted in 81% higher two-dimensional-projected velocities (over 100 cm/s) along the contralateral wall of the aorta. Both graft configurations suggest elevated accumulated shear stresses along individual trajectories. Compared with CFD simulations, HSA successfully characterized the fast-moving flow and hemodynamics in each LVAD graft configuration in vitro , demonstrating the potential utility of this technology as a quantitative imaging modality.
No takes yet. Share an insight, caveat, or question.
Shields et al. (2023) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: