We present a novel noncontact optical method for absolute measurement of refractive index and thickness of optically transparent media. The method is based on a simple dual-confocal fiber-optic sensor design. It includes two independent confocal channels consisting of two identical apertureless fiber-optic-type confocal microscopes constructed by use of a single 2x2 fiber coupler. A geometrical-ray model is used to obtain the analytical dependence between the sample's refractive index and its thickness. The measurement method provides high accuracy in spatially locating the specific imaging points that correspond to the backreflected intensity peaks of the confocal responses. Thus, a simultaneous measurement of the sample refractive index and thickness is achieved.
No takes yet. Share an insight, caveat, or question.
Ilev et al. (2002) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: