Answers are grounded in this paper and cite their sources.
We have examined the sensitivity of sequential two photon laser induced fluorescence (LIF) detection of elemental mercury, Hg(0) in the gas phase. The most sensitive approach involves an initial laser excitation of the 6(3)P1-6(1)S0 transition at 253.7 nm, followed by excitation with a second laser to the 7(1)S0 level. Blue shifted fluorescence is observed on the 6(1)P1-6(1)S0 transition at 184.9 nm. The excitation scheme, involving sequential excitation of two atomic transitions, followed by detection of the emission from a third is extremely specific and precludes detection of anything other than atomic mercury. Using our 10 Hz laser system we have achieved a detection sensitivity of 0.1 ng m(-3) at a sampling rate of 0.1 Hz, i.e. averaging 100 laser shots at a pressure of one atmosphere in air. At low concentrations we sampled simultaneously with an automated mercury analyzer (Tekran 2537A), to ensure accuracy. We have examined the linearity of the technique, generating flows containing mercury concentrations between 1 and 10,000 ng m(-3) using a permeation tube and dynamic dilution, but relying on the concentrations given by the Tekran at low levels and the concentration calculated from dilution at high levels. We find that the detection is linear over the five orders of magnitude that we were able to vary the concentration. Our measured detection limits in He and Ar are much lower as these gases are inefficient fluorescence quenchers.
No takes yet. Share an insight, caveat, or question.
Bauer et al. (2002) studied this question.