Why the study?
Access to ICG vital signs from the upper-arm brachial artery could enable user-convenient wearable armbands for ambulatory long-term monitoring as an alternative to conventional thorax recordings.
Does upper arm impedance cardiography with recursive signal averaging and multi-stage wavelet denoising provide comparable signal quality to thorax impedance cardiography in healthy subjects?
Does upper arm impedance cardiography with recursive signal averaging and multi-stage wavelet denoising provide comparable signal quality to thorax impedance cardiography in healthy subjects?
A novel denoising strategy for upper-arm impedance cardiography yields signal quality and stroke volume estimates highly correlated with standard thoracic measurements, demonstrating potential for long-term wearable hemodynamic monitoring.
Answers are grounded in this paper and cite their sources.
Arm-ICG with recursive averaging may enable wearable hemodynamics in healthy subjects; hypothesis-generating, requires patient validation before practice change.
Impedance cardiography (ICG) is a low-cost, non-invasive technique that enables the clinical assessment of haemodynamic parameters, such as cardiac output and stroke volume (SV). Conventional ICG recordings are taken from the patient’s thorax. However, access to ICG vital signs from the upper-arm brachial artery (as an associated surrogate) can enable user-convenient wearable armband sensor devices to provide an attractive option for gathering ICG trend-based indicators of general health, which offers particular advantages in ambulatory long-term monitoring settings. This study considered the upper arm ICG and control Thorax-ICG recordings data from 15 healthy subject cases. A prefiltering stage included a third-order Savitzky–Golay finite impulse response (FIR) filter, which was applied to the raw ICG signals. Then, a multi-stage wavelet-based denoising strategy on a beat-by-beat (BbyB) basis, which was supported by a recursive signal-averaging optimal thresholding adaptation algorithm for Arm-ICG signals, was investigated for robust signal quality enhancement. The performance of the BbyB ICG denoising was evaluated for each case using a 700 ms frame centred on the heartbeat ICG pulse. This frame was extracted from a 600-beat ensemble signal-averaged ICG and was used as the noiseless signal reference vector (gold standard frame). Furthermore, in each subject case, enhanced Arm-ICG and Thorax-ICG above a threshold of correlation of 0.95 with the noiseless vector enabled the analysis of beat inclusion rate (BIR%), yielding an average of 80.9% for Arm-ICG and 100% for Thorax-ICG, and BbyB values of the ICG waveform feature metrics A, B, C and VET accuracy and precision, yielding respective error rates (ER%) of 0.83%, 11.1%, 3.99% and 5.2% for Arm-IG, and 0.41%, 3.82%, 1.66% and 1.25% for Thorax-ICG, respectively. Hence, the functional relationship between ICG metrics within and between the arm and thorax recording modes could be characterised and the linear regression (Arm-ICG vs. Thorax-ICG) trends could be analysed. Overall, it was found in this study that recursive averaging, set with a 36 ICG beats buffer size, was the best Arm-ICG BbyB denoising process, with an average of less than 3.3% in the Arm-ICG time metrics error rate. It was also found that the arm SV versus thorax SV had a linear regression coefficient of determination (R2) of 0.84.
No takes yet. Share an insight, caveat, or question.
Escalona et al. (2023) studied this question.