Why the study?
What are the cineradiographic functional characteristics of the Björk-Shiley monostrut prosthesis 9 months after implantation?
What are the cineradiographic functional characteristics of the Björk-Shiley monostrut prosthesis 9 months after implantation?
High-speed cineradiography demonstrates consistent functional parameters, including a 70-degree opening angle and measurable disc rotation and slide, for the Björk-Shiley monostrut prosthesis at 9 months post-implantation.
Offers normative cineradiographic parameters for Björk-Shiley monostrut valves; leaves open their role in dysfunction detection pending outcome studies.
In 46 patients, 51 Björk-Shiley monostrut valves were studied with high-speed cineradiography 9 months after implantation. Each valve was viewed in lateral, frontal and oblique projections for 10-12 heart beats. In the aortic valves the opening time was 0.04-0.067 (mean 0.052) and closing time 0.04-0.067 (mean 0.046) sec. In all these valves the maximum opening angle was 70 degrees. Disc rotation could be monitored in 29/31 valves, and during 10 heart beats averaged 20 degrees. Disc slide of 2.0-2.5 mm was seen in 27/31 valves. It occurred during the last 40-50 degrees of opening and the first 20-40 degrees of closing. In mitral valves with sinus rhythm (12/20), the opening time was 0.053-0.106 (mean 0.073) and closing time 0.040-0.106 (mean 0.061) sec. Mean disc rotation (recorded in 17/20 valves) during 10 heart beats was 25 degrees, and disc slide of 2.0-2.5 mm was found in all but one case, mainly during the last 50 degrees of opening and the first 30 degrees of closing. The valve opened to 70 degrees in all but one patient, in whom the disc clearly impinged against the left ventricular wall and opened only to 35-50 degrees at rest. At reexamination a year later the patient was asymptomatic, in sinus rhythm, and the disc had regained full movement with 70 degrees opening.
No takes yet. Share an insight, caveat, or question.
Albrechtsson et al. (1987) studied this question.