Answers are grounded in this paper and cite their sources.
IrO x electrodes suffer from large potential drift in solutions in the presence of strong reductive anions such as S 2− , I − , etc. resulting in the inaccuracy of pH measurement. Our homemade IrO x electrode developed by cycling heat-treatment and quenching process for three times with excellent response property showed the same problem. In-situ/ex-situ XAFS study showed the composition of the IrO x film changes in strong reductive solution. At the meantime, XPS survey of the IrO x electrode also verified the composition change of Ir 4+ /Ir 3+ before and after its immersion in these solutions. In order to solve the above problem, the porous-structured IrO x electrodes were modified by being dipped into Nafion solution for 1∼3 times. Surface and cross section characterizations, EIS tests as well as the pH detection property evaluations were conducted to identify the modification effect and determine a better modification process. Nafion modification for just once is suggested to prevent the electrode from potential drift in solutions containing strong reductive anions. Finally, the pH response mechanism of the modified IrO x electrode in reductive solutions was discussed.
No takes yet. Share an insight, caveat, or question.
Huang et al. (2018) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: