An electronic polarimeter is described in which a signal-sampling technique and feedback are used together to measure periodic retardation changes. A Faraday cell is arranged to feed back a rotation just sufficient to cancel that caused by the unknown retardation, so that the Faraday cell current provides the quantitative measure desired. The photomultiplier output is fed to a sampling oscilloscope in the feedback loop. The instrument is best suited for the measurement of small retardation changes, of the order of 50 mrad or less. The 6-nsec time resolution obtained was limited by the photomultiplier response; a faster detector could improve it considerably. The repeatability obtained of 10(-4) rad was determined by the usual compromise between noise, primarily photon noise, and averaging time. However, other factors limit the accuracy attainable to, at best, +/-3%, for a retardation change of ~20 mrad, while a more typical accuracy was +/-5%. Either Faraday rotation, or small rotations of the optical ellipsoid of aspecimen, could also be observed.
No takes yet. Share an insight, caveat, or question.
A. R. Johnston (1969) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: