Answers are grounded in this paper and cite their sources.
An absolute determination of the yield of the ferrous sulfate radiation dosimeter has been made by delivering to the solution a measured charge of cathode rays at known energies of 1 or 2 Mev. Appropriate small corrections have been made for backscattering, window absorption and, bremsstrahlung. The result, G = 15.6±0.5, is in complete agreement with Hochanadel's calorimetric determination of the yield, but disagrees by 30 percent with the most widely used value obtained by cavity ion chamber comparison.
No takes yet. Share an insight, caveat, or question.
Saldick et al. (1954) studied this question.