Answers are grounded in this paper and cite their sources.
We report the operation of an aerosol analyzer capable of measuring the fluorescence spectra of single micrometer-sized bioaerosol particles as they flow through the instrument. Particles entrained in an airstream initially traverse a cw (continuous wave) (488 nm) “trigger” laser beam where their elastic scattering and total fluorescence is measured with photomultipliers. When the elastic scattering and/or fluorescence signals exceed preset levels, a 266 nm UV “probe” laser is triggered to fire and illuminate preselected particles downstream from the trigger laser. The UV laser-excited spectra of particles are measured with the instrument's image-intensified CCD detector that is gated to be on when the probe laser fires. We demonstrate the ability of the instrument to capture the fluorescence spectra of single micrometer-sized airborne biological particles. Such spectra should be useful in differentiating between biological and nonbiological aerosols and in partially characterizing airborne biological particles.
No takes yet. Share an insight, caveat, or question.
Pinnick et al. (1998) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: