An ultrasensitive UV-visible spectrophotometer, of a single-beam type provided with a photon counting system having a time resolution of 10 ns has been constructed. It has been shown that absorption spectra can be measured to within an error of 10−5 in optical density with this instrument. Noise levels, measured as a function of the photon count numbers, are found to agree with the theoretically expected statistical fluctuation of the photons. Absorption spectra of thin naphthacene films, 32 nra or below in thicknesses, deposited on a quartz plate, were obtained with this instrument, and found to change with the film thickness.
No takes yet. Share an insight, caveat, or question.
Yamamoto et al. (1979) studied this question.