Key result
The FIR highpass filter with a cut-off frequency of 0.67 Hz achieved the best performance for baseline wander removal in real ECG signals.
Why the study?
Which baseline wander removal method provides the best performance for ECG signals?
Which baseline wander removal method provides the best performance for ECG signals?
Answers are grounded in this paper and cite their sources.
A FIR high-pass filter with a cut-off frequency of 0.67 Hz provides the best performance for removing baseline wander from ECG signals compared to other common methods.
Cardiovascular diseases are the leading cause of death worldwide, accounting for 17.3 million deaths per year. The electrocardiogram (ECG) is a non-invasive technique widely used for the detection of cardiac diseases. To increase diagnostic sensitivity, ECG is acquired during exercise stress tests or in an ambulatory way. Under these acquisition conditions, the ECG is strongly affected by some types of noise, mainly by baseline wander (BLW). In this work were implemented nine methods widely used for the elimination of BLW, which are: interpolation using cubic splines, FIR filter, IIR filter, least mean square adaptive filtering, moving-average filter, independent component analysis, interpolation and successive subtraction of median values in RR interval, empirical mode decomposition and wavelet filtering. For the quantitative evaluation, the following similarity metrics were used: absolute maximum distance, the sum of squares of distances and percentage root-mean-square difference. Several experiments were performed using synthetic ECG signals generated by ECGSYM software, real ECG signals from QT Database, artificial BLW generated by software and real BLW from the Noise Stress Test Database. The best results were obtained by the method based on FIR high-pass filter with a cut-off frequency of 0.67 Hz.
No takes yet. Share an insight, caveat, or question.
Romero et al. (2018) studied Baseline wander in ECG signals. FIR highpass filter (cut-off 0.67 Hz) vs. Other baseline wander removal methods (IIR, Splines, LMS, MAF, ICA, ISSM, EMD, Wavelet) was evaluated on Signal similarity metrics (Absolute maximum distance [MAD], sum of squares of distances [SSD], percentage root-mean-square difference [PRD]). The FIR highpass filter with a cut-off frequency of 0.67 Hz achieved the best performance for baseline wander removal in real ECG signals.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: