The incorporation of an aromatic hydrocarbon fluorescent compound into silicone allows utilization of fluorescence quenching as a mechanism to measure molecular oxygen concentration in the composite matrix. Excitation and emission spectra of the aromatics change in agreement with the Franck-Condon model when the universal solvent is benzene. Fluorescence intensity was measured as a function of fluorophor concentration in several solvents: benzene; silicone fluids of different viscosities; and fully polymerized silicone. The relative fluorescence intensity of 9,10-diphenyl anthracene in poly(dimethyl siloxane) and silicone fluids was measured as a function of oxygen concentration, fluorophor concentration, and fluid viscosity. This results in a linear relationship between relative fluorescence intensity and concentration of oxygen.
No takes yet. Share an insight, caveat, or question.
Cox et al. (1985) studied this question.