The use of the hemispherical deflection analyser (HDA) is well established in electron spectroscopy. Increasing use has been made of two HDAs placed in tandem. Such an arrangement has useful advantages over a single HDA. Usually the two analysers are placed together without an intervening lens system so that the exit plane of the first is placed in close proximity to the entrance plane of the second. When two HDAs are placed together in this way consideration must be given to the correction of the fringing fields of the analysers where they meet. This article describes one solution to this problem that should prove especially useful for the case where the tandem arrangement is used on conjunction with a multichannel detection system, an arrangement that is expected to find increasing use.