Answers are grounded in this paper and cite their sources.
A new method was developed and described for the measurement of airborne formaldehyde using derivatization with 3,5-bis(trifluoromethyl)phenylhydrazine (TFMPH) coated onto silica solid phase extraction cartridges. Analysis by GC-ECD provides a detection limit of 74 ng formaldehyde per sample. A field study was conducted to compare the use of TFMPH to 2,4-dinitrophenylhydrazine (DNPH) and NIOSH method 3500 (chromotropic acid, CTA). Samples were collected from indoor and outdoor environments known or suspected to contain formaldehyde. Use of TFMPH with GC-ECD analysis correlates well with both methods (R2=0.93, slope=1.07 vs. DNPH; R2=0.99, slope=1.06 vs. CTA). Spiked samples were shown to be stable at least 7 days when stored at –20 °C. Analysis of samples by GC-MS with selected ion monitoring (GC-MS/SIM) also proved feasible. Laboratory and field results show the use of TFMPH to be viable for quantifying airborne formaldehyde in occupational and environmental samples.
No takes yet. Share an insight, caveat, or question.
Marsella et al. (2000) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: