Answers are grounded in this paper and cite their sources.
A theoretical analysis of internal heat transfer processes and their variation with temperature over the range 4.2–100 K is made for a typical commercial germanium resistance thermometer. It is shown that the proportion of the heat transfer occurring through the solid materials in the thermometer compared to that taking place in the gas trapped in the thermometer capsule varies substantially with temperature. The factors influencing heat transfer include the construction and the size of the thermometer, the thermal properties of materials, and the position of the germanium crystal within its capsule. Where heat transfer through the gas in the capsule is important, any changes in the composition of the gas play a significant role through the consequent changes in its thermal conductivity.
No takes yet. Share an insight, caveat, or question.
Hai et al. (1993) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: