The thermodynamic temperature scale was realized over the range from 0 °C to -183 °C by using a helium constant volume gas thermometer, and platinum resistance thermometers were calibrated to derive an accurate relation between the resistance of pure platinum and the thermodynamic temperature. It is shown that the International Practical Scale of Temperature, as defined by the platinum resistance thermometer using the Callendar-Van Dusen equation, is in error by -0.036 °C at -80 °C and +0.023 °C at -150 °C, the estimated limit of error on these differences being ±0.006 °C.